Meaning
Medical device is the regulated class of products a printed part may be meant for, governed in the EU by Regulation (EU) 2017/745. The powder and resin sheets in these resources either disclaim that use or list it without a declaration. A claim that a part suits such use is a written declaration of the part’s maker. The material’s datasheet is not that declaration.Powder Sheets
The EOS PA 2200 sheet states that the powder is not developed or certified as a medical device under Regulation (EU) 2017/745. By that sheet, responsibility for any medical use rests with the user. The EOS process data sheet for 316L-4404 carries the same statement and adds the older Directive 93/42/EEC.Read More
Chemistry and Approval
The EOS 316L material data sheet lists medical devices among typical applications, with no declaration. It states that the powder’s composition corresponds to ASTM F138, the wrought implant steel UNS S31673. That is a statement about chemistry. It does not make a printed part an implant, and it does not stand in for the maker’s conformity process.Resins
Formlabs handles its biocompatible resins apart in the wash. It points to each resin’s own instructions, calls for isopropyl alcohol (IPA) of 99 % or higher and asks for separate wash buckets. Whether a washed and cured part suits a medical use remains a declaration of the resin’s maker and the part’s maker.Written Declaration
Annex A of ISO/ASTM 52901 gives the typical content of a purchase order. A part with a regulated use or a safety function needs more than the typical. An order asks in writing for the declarations of the part’s maker, for the exact grade and build parameters.